Metode si echipamente folosite

Spectrometrie gama de inalta rezolutie
  • Instalatie de masurare a activitatii gama cu sistem spectrometric HPGE cu analizor multicanal CANBERRA +SILENA
  • Caracteristici:
    Detector HPGE – Silena model RGC-P1519
    Preamplificator incorporat (racit) – model RFP 110;
    Incinta ecran+colimator de Pb
    AMC – CANBERA – INSPECTOR  tip 1200, cu HV Incorporata
    Interval energetic 50 keV – 2049 keV
    Interfata calculator TI-Pentium-LAPTOP
    Software GENIE-2000;
    GAMMA ANALISYS , SPECTROSCOPY ASSISTANT V.2.0/2001;
  • Instalatie de masurare a activitatii gama cu sistem spectrometric cu analizor multicanal HPGE
  • Caracteristici:
    Detector HPGE – Canberra model GC-2518
    Preamplificator incorporat(racit) – model CSL-2002 110;
    Incinta ecran de Pb
    AMC – CANBERA – INSPECTOR  tip 1270, cu HV Incorporata
    Interval energetic 30 keV – 1718 keV
    Interfata calculator DTK-Pentium
    Software GENIE-2000;
    GAMMA ANALISYS , SPECTROSCOPY ASSISTANT V.2.0/2001;

  • Sistem mobil de caracterizare a materialelor radioactive ISO-Cart Integration and Test Record ORTEC
  • Caracteristici:
    Detector HPGE;
    Carucior versatil ISO-Cart
    Incinta ecran+colimatoare de Pb si W
    Platforma rotitoare pentru containere cu masa maxima de 450 kg
    Dispozitiv manipulare containere si cantarire
    AMC Model Digi Dart
    Interval energetic 60 keV – 1840 keV
    Calculator LAPTOP Dell Model Latitude D520
    Software ISOTOPIC - 32; Gamma Vision

Spectrometrie beta cu scintilatori lichizi
  • Instalatie de spectrometrie cu scintilatori lichizi de tip TRI-CARB 2100 TR
  • Caracteristici:
    Detectori cu 2 tuburi fotomultiplicatoare
    Domeniu maxim energetic: 2000 keV
    Domenii prestabilite:
    Tritiu: 0-18.6 keV
    Carbon 14: 0-156 keV
    Fosfor 32: 5-1700 keV
    Iod 125: 0-70 keV
    Eficienta minima acceptabila:
    Tritiu: 60%
    Carbon 14:95%

Aparatura portabila pentru masurarea directa a campurilor de radiaþii
  • Radiometru/spectrometru portabil, model InSpector 1000 CANBERRA
  • Caracteristici:
    Detector NaI 1,5 ´ 1,5cc – interval energetic 50 keV – 3 MeV
    Detector GM incorporat – interval energetic 30 keV – 1.4 MeV
    Nuclid Identification (NID)  Mode GENIE-2000
    Capacitatea de stocare >256 spectre
    USB HOST-USB DEVICE

  • Radiometru/contaminometru portabil, model FH40 GL10
  • Caracteristici:
    Detector GM incorporat
    Sonda FAG model FHZ 120 A
    Sonda FAG model FHZ 140
    Interval energetic 45 keV – 1.3 MeV
    Domeniul de masura 0.01 µSv/h – 9.99mSv/h

Radiometrie
  • Radiometru alfa/beta cu fond scazut, model MPC 9300, Protean Instrument Corporation
  • Caracteristici:
    Contori proporþionali cu circulaþie de gaz (P-10 Argon-Metan)
    Fereastrã de 80 mg/cm2 - fond ultrascãzut
    Eficienþa pentru 90Sr-90Y 55%,
    Eficienþa pentru 241Am 42%.
    Voltage Range  705V – 1875 V
    Calculator compatibil MS-DOS/MS Windows
    PICNET Software

Dr. Alexandru Toma – sef de laborator
Tel.:
+40 248 213400 ext. 524
Fax.: +40 248 262449
e-mail: alexandru.toma@nuclear.ro

Dr. Cristian Dulama – responsabil tehnic
Tel.:
+40 248 213400 ext. 511
Fax.: +40 248 262449
e-mail: cristian.dulama@nuclear.ro

Mr. Hirica Ovidiu – responsabil asigurarea calitatii
Tel.:
+40 248 213400 ext. 109
Fax.: +40 248 262449
e-mail: ovidiu.hirica@nuclear.ro

Mr. Relu Dobrin
Tel.:
+40 248 213400 ext. 511
Fax.: +40 248 262449
e-mail: relu.dobrin@nuclear.ro


Recomandari: Rezolutie video
minim 1024 x 768,
Internet Explorer 6

Webmaster: webmaster@nuclear.ro


Copyright © 2009 S.C.N. Pitesti
sus